Description
The Kanomax 6006-DC Anemomaster is a compact, high-precision thermal anemometer engineered for critical low-velocity air measurements. Driven by the Original Engineering of Kanomax hot-wire technology, this diagnostic tool provides simultaneous air velocity and temperature feedback. This variant includes a dedicated DC power and analog output module, ensuring continuous laboratory or cleanroom monitoring capabilities without battery depletion risk.
SYSTEM PARAMETERS AND SPECIFICATIONS
Manufacturer: Kanomax USA, Inc.
Model: 6006-DC (Anemomaster Lite with DC Output)
Sensor Type: Platinum-coated hot-wire and thermistor
Air Velocity Measurement Range: 0.10 to 20.0 m/s (20 to 3940 FPM)
Air Velocity Measurement Accuracy: ±5.0% of reading or ±0.015 m/s (whichever is larger)
Air Velocity Resolution: 0.01 m/s (0.10 to 9.99 m/s), 0.1 m/s (10.0 to 20.0 m/s)
Temperature Measurement Range: 0 to 50 °C (32 to 122 °F)
Temperature Measurement Accuracy: ±1.0 °C (±1.8 °F)
Temperature Resolution: 0.1 °C
Analog Output Interface: DC 0 to 1 V (scaled to air velocity)
External Power Input: 5 to 6 VDC
Internal Power Source: 4 x AA Alkaline batteries
Battery Operation Lifetime: Approx. 8 hours (at 20 °C, air velocity 5.0 m/s)
Probe Diameter at Tip: Ø 6.1 mm
Probe Base Diameter: Ø 10.3 mm
Probe Extension Length: 150 mm to 960 mm
Cable Length: 1500 mm
Display Type: Liquid Crystal Display (LCD) with backlight
Operating Temperature Range: 5 to 40 °C (no condensation)
Storage Temperature Range: -10 to 50 °C (no condensation)
Weight: 180 grams (excluding batteries)
Certifications: NIST Traceable, CE, RoHS
INDUSTRIAL APPLICATION SCENARIO
In pharmaceutical cleanrooms and semiconductor processing bays, laminar airflow integrity is paramount to prevent particulate contamination. Facility engineers deploy the Kanomax 6006-DC to verify that unidirectional flow systems maintain the standard 0.45 m/s (90 FPM) velocity vector. Its slender telescopic probe can be inserted through narrow ceiling grilles, eliminating the need to dismantle protective HEPA filtration housings. The 0 to 1 V DC analog output allows Technical Compatibility with standard data loggers, enabling continuous spatial flow profile mapping over 24-hour cycles to identify turbulent zones.
PROCUREMENT AND LIFECYCLE MANAGEMENT
Original calibration certificates traceable to NIST are provided with every instrument to guarantee standard compliance from day one. Please contact our distribution department to confirm current Lead Time and availability for high-volume deployments or calibrated spare sets.
Key Attributes
Resolution: 0.01m/s
Temperature Range: -4 to 158°F (-20 to 70°C)
Type: Anemometer
Place of Origin: Japan
Material: Plastic
Starting Wind Speed: 0.01 to 30.48m/s
Wind Speed Unit Selection: OBM
Packaging: Plastic Box
Model Number: 6006-DC
Brand Name: Kanomax
Velocity Measurement Range: 0.01 to 20 m/s (2 to 3,940 FPM)
Velocity Resolution: 0.1 m/sec (10 to 20 m/s)
Velocity Accuracy: ±0.015 m/s
Power Source: Battery powered
Operating Temperature: 41 to 104°F (5 to 40°C)
Approvals: CE
Battery Type: AA
Packaging and Delivery
Selling Units: Single item
Single package size: 37X31X14 cm
Single gross weight: 1.800 kg











Reviews
There are no reviews yet.